HomeMy WebLinkAboutE603 Enlarged Mammography PlansFJFJMAMMOGRAPHY42715ACACSBP-4A/1REMOTE EMERGENCY OFFS.T. SWITCHWARNING LIGHTPROVIDE 3-#8 & 1-#10 E.G. IN 3/4" C 21CB1FD2WV2WH13WV1FD115MAMMOGRAPHY(STEREOTATICBIOPSY)431ACPP4B/19ACWARNING LIGHTWARNING LIGHTREMOTE EMERGENCY OFF S.T. SWITCHCB1WV1WV2WH1FD1FD2123PROVIDE 3-#8 & 1-#10 E.G. IN 3/4" CDESCRIPTION:WARNING LIGHT = 4" STANDARD BOX WITH "X-RAY ON"OR RED WARNING LAMP OVER DOOR. USE APPROVED FIXTURES - INCANDESCENTOR LED, REFER TO LOCAL CODES.(SUPPLIED & INSTALLED BY HOSPITAL'S ELECTRICAL CONTRACTOR.)HOSPITAL'S ELECTRICAL CONTRACTOR TO PROVIDE POWER TO THE LAMP, ANDA SWITCH LEG TO THE "SELENIA DIMENSIONS" GANTRY FOR CONTACT CLOSUREONLY. AN XRAY EXPOSURE WILL CLOSE CONTACT, ENERGIZING THE LIGHT.THE CONTACT RELAYS ARE RATED AT: 250VAC, 10 AMP NORMALLY OPEN,AND 10 AMP, 30VDC NORMALLY OPEN.OPTIONALDOORSWITCH = MICROSWITCH IN DOOR JAMB TO ENABLE "ROOM IN USE"LIGHT OR "X-RAY SYSTEM POWER ON" INDICATOR.NOTE: CUSTOMER MAY USE A WALL SWITCH FOR THIS PURPOSE IF DESIRED.(TYPICALLY USED ONLY WHERE REQUIRED BY CODE, OR CUSTOMER PREFERENCE)TYPICAL DOOR SWITCH/WARNING LIGHT DETAILFINISH FLOORLINEALTERNATE SIDEDOORSWITCHTWO #14 FORDOORSWITCH*CONTRACTOR TO PROVIDE 120V POWER TO LIGHT &RUN A SWITCH LEG INTO "SELENIA DIMENSIONS" GANTRYWIRE SIZES:TWO # 14 + GROUND TOWARNING LIGHT*FINISH CEILING LINE(WL) WARNING LIGHTUSE APPROVED FIXTURES-INCANDESCENT OR LEDREFER TO LOCAL CODES1/2"CONDUIT FROM WARNING LIGHT TODOORSWITCH(DS) OPTIONAL DOORSWITCHLOCATE DOORSWITCH ON HINGE SIDE OF DOOR AS REQ'D BY ANY LOCAL CODES3/4" CONDUIT TO WIREWAY OR "JB1" NEAR " SELENIA DIMENSIONS" GANTRYTYPICAL CIRCUIT BREAKER DIAGRAM*CUSTOMER MAY HAVE TO STEP-UP EXISTING INCOMING POWER WITHTRANSFORMER TO MEET HOLOGIC'S POWER SPECIFICATIONS.MUST MEET ALL ELECTRICAL SPECIFICATIONS.PLEASE VALIDATE PRIOR TO INSTALLATION, AND/OR CONSULTWITH LOCAL HOLOGIC FIELD ENGINEER OR PROTECT MANAGER PRIOR TO INSTALLATION. IF STEP-UP TRANSFORMER IS USED FOR THIS APPLICATIONTRANSFORMER MUST HAVE LESS THAN .15 OHMS OF IMPEDANCE.TO MAIN DISTRIBUTION PANELSIZE PER CODESIZE PERCODEREMOTE EMERGENCYOFF SHUNT TRIP SWITCH(IF REQUIRED)CONSULT RADIOLOGY PERSONNELCONCERNING LOCATION.OR LOCATE PER LOCAL CODES.CIRCUIT BREAKER(CB1)TO HORIZONTAL WIREWAY 'WH1'PROVIDE (2) CONDUCTOR #8 WIRES AND (1) #8 AWG GROUND WIRE WITH 36" NON -TERMINATED PIGTAILS EXTENDING FROM 'FD1'. THE SELENIA DIMENSIONS GANTRY IS TO BE PERMANENTLY WIRED.5'-0" [152cm] TO FINISH FLOOR.(OR PER LOCAL CODES)5'-0" [152cm] TO FINISH FLOOR.(OR PER LOCAL CODES)803573SNSEWYORKOFETAT SSOIANLEOFPRLI C E N S E DREENIGN ETRASS N EJ.CIRER EXCELSIOR°°°°°°°°DATE:PROJECT:OTHER:DRAWN BY:ABCDEFGABCDEFG1234567891012345678910ArchitecturePlanningInterior Design1112HI1112HI619 West State StreetIthaca, New York 14850132 E. Jefferson StreetSyracuse, New York 13202p 607 273 7600 www.holt.com2/10/2021 1:59:38 PME603ENLARGEDMAMMOGRAPHYPLANS02/05/20212018089AuthorCayuga Park, Ithaca, New York 14850Park Grove - Cayuga Medical CenterMEDICAL OFFICE BUILDING1/2" = 1'-0"1ENLARGED MAMMOGRAPHY ROOM1/2" = 1'-0"2ENLARGED MAMMOGRAPHY(STEREROTATIC BIOPSY) PLAN3WARNING LIGHT DETAIL4CIRCUIT BREAKER DIAGRAMCB1DEDICATED CIRCUIT BREAKER/DISCONNECT WITH REMOTE EMERGENCY SHUNT TRIP.(SHUNT TRIP USED ONLY WHERE REQUIRED BY LOCAL CODE OR CUSTOMER)SUPPLY 200-240 VAC, 50-60 HZ, SINGLE PHASE, 40 AMP, HACR, OR UL 489 LISTEDCIRCUIT BREAKER/DISCONNECT.PROVIDE (2) CONDUCTOR #8 AWG WIRES, AND (1) #8 AWG GROUND WIRE FROM 'CB1' TO 'WH1' WITH 36" NON-TERMINATED PIGTAIL EXTENDING FROM THE END OF 'FD1'. PROVIDE CONDUIT TO 'WH1'. SIZED PER CODE. MAXIMUM LINE IMPEDANCE NOT TO EXCEED: .20 OHMS FOR 208 VAC (IEC 601-2-7, TABLE 101). SEE ELECTRICAL SPECIFICATIONS. LOCATE PER LOCAL CODE. GANTRY TO BE PERMANENTLY WIRED. THE CONNECTION OF CUSTOMER SUPPLIED POWER WIRES TO THE GANTRY IS TO BE PERFORMED BY CUSTOMER'S LICENSES ELECTRICIAN.CO1 2" DIA. GROMMETED CUTOUT IN THE END FITTING, OR COVER OF 'WH1'WH1 HORIZONTAL SURFACE RACEWAY, 1 3/4" X4 3/4" WIREMOLD 4000 SERIES OR EQUAL, DIVIDED, WITH REMOVABLE COVERS, BOTTOM OF WIREWAY AT 12" AFF. PROVIDE FULL ACCESS TO 'WV1', AND 'WV2'WV1,WV2 VERTICAL SURFACE WIREWAY, 1 3/4" X 4 3/4" WIREMOLD 400 SERIES OR EQUAL, DIVIDED, WITH REMOVABLE COVERS, PROVIDE FULL ACCESS TO 'WH1', AND CORRESPONDING FLOOR DUCT (WV1 TO FD1/WV2 TO FD2).FD1HORIZONTAL SURFACE RACEWAY, 1 3/4" X4 3/4" WIREMOLD 4000 SERIES OR EQUAL, DIVIDED, WITH REMOVABLE COVERS, PROVIDE FULL ACCESS TO 'WV1'. PROVIDE 3/4" FLEXIBLE METALLIC CONDUIT (SEALTITE OR EQUAL) TO GANTRY, AND GROMMETED OPEN END.FD2 HORIZONTAL SURFACE RACEWAY, 1 3/4" X4 3/4" WIREMOLD 4000 SERIES OR EQUAL, DIVIDED, WITH REMOVABLE COVERS, PROVIDE FULL ACCESS TO 'WV2'. PROVIDE GROMMETED OPEN END.FD2 HORIZONTAL SURFACE RACEWAY, 1 3/4" X4 3/4" WIREMOLD 4000 SERIES OR EQUAL, DIVIDED, WITH REMOVABLE COVERS, PROVIDE FULL ACCESS TO 'WV2'. PROVIDE GROMMETED OPEN END.DDEDICATED 120V, 20 AMP HOSPITAL GRADE DUPLEX OUTLET. (AQUISITION WORKSTATION). ** 120V OUTLET AND NETWORK OUTLETS FOR AWS CONSOLE MAY BE INSTALLED IN COVER OF 'FD2' NEAR END120V, 20 AMP DUPLEX OUTLET. (FOR SERVICE/TEST EQUIPMENT)WALL OUTLET BOX FOR CAT 5 CABLE. (HOSPITAL NETWORK CONNECTION). RECOMMEND HOLOGIC CONNECT FOR REMOTE DIAGNOSTIC SERVICE, SWITCH GIGABIT ETHERNET (1000BASE-T)DEDICATED OUTSIDE PHONE LINE (CUSTOMER TO DETERMINE REQUIRMENT/LOCATION)REMOTE EMERGENCY OFF SHUNT TRIP SWITCH, LOCATE PER CODE. SEE DETAIL ON THIS SHEET. 12PROVIDE 3/4" CONDUIT AS REQUIRED TO CIRCUIT BREAKER 'CB1' FROM INCOMING POWER SOURCE.PROVIDE 3/4" CONDUIT FROM 'CB1' TO 'WH1'3PROVIDE 3/4" DIAMETER FLEXIBLE METALLIC CONDUIT (SEALTITE OR EQUAL) FROM END OF 'FD1' TO BACK OF GANTRY. USE ELBOW FITTING IN END OF 'FD1'. LENGTH SHOULD ALLOW FOR A 12-18" SERVICE LOOP.ITEM DESCRIPTIONSUPPLIED AND INSTALLED BY HOSPITAL'S ELECTRICAL CONTRACTORELECTRICAL LEGENDSUPPLEMENTAL NOTESALL ITEMS LISTED BELOW ARE TO BE PROVIDED/DESIGNED BY THE ARCHITECT.THE ARCHITECT SHALL CONSULT WITH THE CUSTOMER FOR ANY ADDITIONAL INFORMATION .1. RECOMMEND INCANDESCENT OR LED, AS WELL AS FLUORESCENT LIGHT FIXTURES.THE INCANDESCENT OR LED FIXTURES SHOULD BE DISTRIBUTED AROUND THE ROOM. HOLOGIC RECOMMENDS AGAINST THE USE OF DIMMER CONTROLS, AS THEY CAN CAUSE ISSUES WITH EMI INTERFERENCE.2. IT IS RECOMMENDED THAT ALL DIAGNOSTIC ROOMS SHOULD HAVE TILE OR VINYL FLOORING.3. A TILE DROP CEILING IS RECOMMENDED FOR ANY DIAGNOSTIC ROOMS.4. ALL CABINETRY WORK IS TO BE DESIGNED AND PROVIDED BY THE ARCHITECT. RECOMMEND ALL COUNTERS HAVE ROUNDED CORNERS.GROUNDING NOTES*FIELD VERIFY ACTUAL APPLICATION WITH LOCAL LICENSED ELECTRICIAN.THE SOURCE GROUND CONNECTION FOR THIS SYSTEM WILL BE THE GROUND SIDE OF THE NEUTRAL TO GROUND BONDING POINT FOR THE POWER DELIVERED TO THIS SYSTEM. THE SOURCE WILL BE EITHER THE MAIN SERVICE ENTRANCE PANEL OR A SEPARATELY DERIVED SOURCE, SUCH AS AN ISOLATION TRANSFORMER. SYSTEM LOCALIZED COLD WATER PIPING OR BUILDING STEEL ARE NOT ACCEPTABLE GROUND SOURCE CONNECTION POINTS. THE SOURCE GROUND CONDUCTOR SHALL BE THE SAME SIZE AND TYPE AS THE PHASE CONDUCTORS SUPPLYING THE SYSTEM. REFER TO LOCAL AND NATIONAL ELECTRIC CODES FOR GROUNDING REQUIREMENTS.WIRING NOTESELECTRICAL CONTRACTOR SHALL FURNISH AND INSTALL ALL CONDUITS, WIRE, WIREWAYS, DISTRIBUTION TRANSFORMER, JUNTION BOXES, ENCLOSURES, CIRCUIT BREAKERS, ETC. ALL WIRES TO BE THHN TYPE, STRANDED COPPER UNLESS OTHERWISE SPECIFIED. THE ELECTRICAL CONTRACTOR SHALL SUPPLY LUG PAIRS OF #8 AWG OR GREATER FOR ALL CABLES LISTED WITHIN THE SCOPE OF WORK. CONDUIT AND DUCT RUNS ARE SCHOWN SCHEMATICALLY. ACTUAL BUILDING CONDITIONS WILL DETERMINE CONDUIT AND DUCT ROUTES. CONDUIT TURNS TO HAVE LARGE SWEEPING BENDS WITH MINIMUM RADIUS AS SPECIFIED IN ART. 346-10. IF LARGE BENDS IN CONDUIT ARE NOT POSSIBLE, USE SERVICE ENTRANCE ELBOW CONNECTORS TO AVOID DAMAGE TO CABLE. ELECTRICAL CONTRACTOR TO PROVIDE 10'-0" TAILES FOR ALL WIRES UNLESS OTHERWISE SPECIFIED AND IDENTIFY BOTH ENDS OF ALL WIRES. ELECTRICAL CONTRACTOR SHALL MAKE EVERY EFFORT TO ENSURE THAT POINT TO POINT CONDUIT RUNS ARE AS SHORT AS FIELD CONDITIONS ALLOW. LEAVE PULL WIRE IN EMPTY CONDUITS. WORK SHALL COMPLY WITH REGULATORY AGENCIES OF JURISDICTION ALL SUCH AGENCIES SUPERSEDE THESE DRAWINGS AND MUST BE ADHERED TO. THE ELECTRICAL CONTRACTOR SHALL BE RESPONSIBLE FOR SCRUTINIZING THESE DRAWINGS AND SHALL MAKE CORRECTIONS ACCORDING TO PREVAILING CODES. NOTE: CONDUITS MAY BE INCREASED IN SIZE BUT NEVER DECREASED WITHOUT CONSENT FROM THE EQUIPMENT REPRESENTATIVES. ELECTRICAL CONTRACTOR SHALL BE RESPONSIBLE FOR INSTALLING THE PATIENT GROUNDING NETWORK PER NATIONAL ELECTRIC CODE ARTICLE 517 AND NFPA 70. ALL CONDUIT SHALL BE RIGID TYPE UNLESS SPECIFIED OTHERWISE IN THIS DRAWING SET. ALL CONDUIT RUNS EXPOSED TO WET ENVIRONMENTS SHALL BE MADE LIQUID TIGHT. (e.g. UNDERGROUND, ETC) ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS SHALL HAVE OVERSIZED CIVER PLATES. ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS EXPOSED TO A WET ENVIRONMENT SHALL BE MADE LIQUID TIGHT (e.g. IN FLOOR, ETC) THE ELECTRICAL CONTRACTOR SHALL INSTALL, PROVIDE POWER TO AND CONNECT THE FOLLOWING IF REQUIRED: THE "X-RAY IN USE" WARNING SIGN; THE 120 VAC, 15 AMP RECEPTACLE; THE DOOR INTERLOCK SWITCH; THE EMERGENCY OFF SWITCH; AND THE 120VAC, DEUPLEX OUTLETS FOR SERVICE AND HOUSEKEEPING.REVISION SCHEDULENAME DATEReceived by DiMarco 2/15/2021