HomeMy WebLinkAboutE602 Enlarged Radiology PlanJJJDIAGNOSTICRADIOGRAPHY416ACSBP-4A/11SBP-4A/121443345WARNING LIGHTWARNING LIGHT42EMERGENCYX-RAY SHUT-OFFELECTRICAL CONTRACTOR TO PROVIDE LEAD LINING AROUND ALL WALL BACK BOXES EQUAL TO THICKNESS PROVIDED IN DRYWALL. * RELAY WITH A 24 VOLT, DC COIL (NOT TO EXCEED 5 WATTS), ANDA STEERING DIODE WIRED OVER THE RELAY COIL, SUPPLIED ANDINSTALLED BY ELECTRICAL CONTRACTOR. SIEMENS TO PROVIDECONTROL SIGNAL TO RELAY. CONTRACTOR SUPPLIED ANDINSTALLED JUNSTION BOX.SIEMENSPOWER/GENRATORCABINETSUPPLIED 24V(FOR DOORSWITCH ONLY)DSSUPPLIED 24VDCLIGHTING PANEL*RELAYJUNCTION BOX6FT. PIGTAILS"X-RAY ON"LEDWARNINGLIGHTLIGHTING PANEL TOBE PROVIDED WITHSWITCH TO TURN OFFWARNING LIGHTS FORSERVICE PURPOSEMAINDISCONNECTPANEL75AGROUND CABLE (PE)MAIN 480 VTHREE PHASE SUPPLYEMERGENCY OFF BUTTONGENERATOR(SYSTEM CABINET)WARNING LIGHT CONTROLWARNING LIGHTDOOR INTERLOCK SWITCHSINGLE PHASE POWER14 BLACK14 WHITE14 GREEN14 BLACK14 WHITE14 GREEN14 BLACK14 WHITE14 GREEN14 BLACK14 WHITE14 GREENBLACK(3)GREEN(1)BLACK(3) WHITE(1)14 WHITE14 RED(2) 14 BLACKWARNING LIGHT14 BLACK14 WHITE14 GREENELECTRICAL CONDUIT LEGENDNO.CONDUIT12345(1)1" C(1)1-1/2" C(1)2" C(1)2-1/2" C(1)3-1/2" C803573SNSEWYORKOFETAT SSOIANLEOFPRLI C E N S E DREENIGN ETRASS N EJ.CIRER EXCELSIOR°°°°°°°°DATE:PROJECT:OTHER:DRAWN BY:ABCDEFGABCDEFG1234567891012345678910ArchitecturePlanningInterior Design1112HI1112HI619 West State StreetIthaca, New York 14850132 E. Jefferson StreetSyracuse, New York 13202p 607 273 7600 www.holt.com2/10/2021 1:59:34 PME602ENLARGEDRADIOGRAPHYPLAN02/05/20212018089AuthorCayuga Park, Ithaca, New York 14850Park Grove - Cayuga Medical CenterMEDICAL OFFICE BUILDING1/2" = 1'-0"1ENLARGED DIAGNOSTIC RADIOGRAPHY PLANGENERAL NOTES:A. ALL WIRES SPECIFIED SHALL BE COPPER STRANDED, FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR STUBBED CONDUIT ENDS. ALL CONDUCTORS, POWER, SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR DUCT SYSTEM. ELECTRICAL CONTRACTOR SHALL RING OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS MUST BECONTINUOUS COPPER STRANDED AND FREE FROM SPLICES.B. ALUMINUM OR SOLID WIRES ARE NOT ALLOWED.C. WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER SIZES MAY BE REQUIRED BY LOCAL CODES.D. IT IS RECOMMENDED THAT ALL WIRES BE COLOR CODED, AS REQUIRED IN ACCORDANCE WITH NATIONAL AND LOCAL ELECTRICAL CODES.E. CONDUIT SIZES SHALL BE VERIFIED BY THE ARCHITECT, ELECTRICAL ENGINEER OR CONTRACTOR, IN ACCORDANCE WITH LOCAL OR NATIONAL CODES.F. CONVENIENCE OUTLETS ARE NOT ILLUSTRATED. THEIR NUMBER AND LOCATION ARE TO BE SPECIFIED BY OTHERS. LOCATE AT LEAST ONE CONVENIENCE OUTLET CLOSE TO THE SYSTEM CONTROL, THE POWER DISTRITBUTION UNIT AND ONE ON EACH WALL OF THE PROCEDURE ROOM. USE HOSPITAL APPROVED OUTLET OR EQUIVALENT.G. GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED. CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEAT FROM OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO CEILING MOUNTING COMPONENTS AND WIRING IF HIGH WATTAGE BULBS ARE USED. RECOMMEND LOW WATTAGE BULBS NO HIGHER THAN 75 WATTS AND USE DIMMER CONTROLS (EXCEPT MR). DO NOT MOUNT LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING MOUNTED ACCESSORIES WILL BE PARKED.H. ROUTING OF CABLE DUCTWORK, CONDUITS, ETC., MUST RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE NEED FOR GREATER THAN STANDARD CABLE LENGTHS (REFER TO THE INTERCONNECTION DIAGRAM FOR MAXIMUM USABLE LENGTHS POINT TO POINT).I. CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND LOCAL ELECTRICAL CODES.J. A SPECIAL GROUNDING SYSTEM IS REQUIRED IN ALL PROCEDURE ROOMS BYSOME NATIONAL AND LOCAL CODES. IT IS RECOMMENDED IN AREAS WHERE PATIENTS MIGHT BE EXAMINED OR TREATED UNDER PRESENT, FUTURE, OR EMERGENCY CONDITIONS. CONSULT THE GOVERNING ELECTRICAL CODE AND CONFER WITH APPROPRIATE CUSTOMER ADMINISTRATIVE PERSONNEL TO DETERMINE THE AREAS REQUIRING THIS TYPE OF GROUNDING SYSTEM.K. THE MAXIMUM POINT TO POINT DISTANCES ILLUSTRATED ON THIS DRAWINGMUST NOT BE EXCEEDED.L. PHYSICAL CONNECTION OF PRIMARY POWER TO GE EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL CONTRACTOR WITH THE SUPERVISION OF A GE REPRESENTATIVE. THE GE REPRESENTATIVE WOULD BE REQUIRED TO IDENTIFY THE PHYSICAL CONNECTION LOCATION, AND INSURE PROPER HANDLING OF GE EQUIPMENT.M. GEHC CONDUCTS POWER AUDITS TO VERIFY QUALITY OF POWER BEING DELIVERED TO THE SYSTEM. THE CUSTOMER'S ELECTRICAL CONTRACTOR IS REQUIRED TO BE AVAILABLE TO SUPPORT THIS ACTIVITY.N. REFER TO VENDOR DRAWINGS FOR ADDITIONAL INFORMATION. OBTAIN LATEST SET OF VENDOR EQUIPMENT DRAWINGS PRIOR TO STARTING WORK. WHERE DISCREPANCIES EXIST BETWEEN THESE DRAWINGS AND VENDOR DRAWINGS THE VENDOR DRAWINGS SHALL TAKE PRECEDENCE.GENERAL NOTES -CONDUIT ROUTING:A. COORDINATE ALL CONDUIT ROUTES WITH PROJECT MANAGER, THE BUILDINGOWNER AND EXISTING MECHANICAL AND ELECTRICAL EQUIPMENT PRIOR TO INSTALLATION.B. PROVIDE PULL BOXES IN ALL CONDUIT RUNS AS REQUIRED TO FACILITATEPULLING IN CABLES.C. PROVIDE FLOOR AND WALL CORES AS REQUIRED. PROVIDE FIRESTOPPING AT ALL FLOOR AND WALL PENETRATIONS AFTER CABLES HAVE BEEN INSTALLED.2RADIOLOGY WARNING LIGHT & DOOR SWITCH DETAIL3RADIOLOGY POWER DIAGRAMREVISION SCHEDULENAME DATEReceived by DiMarco 2/15/2021